GB/T 27991-2011

Active

Fundamental technical requirements for sediment measurement and particle size analysis instrument in open channel

河流泥沙测验及颗粒分析仪器 基本技术条件

Standard Type
GBT
ICS
07.060
CCS
N 93
Status
Active
Issue Date
2011-12-30
Implementation
2012-06-01
Centralized Committee
全国水文标准化技术委员会水文仪器分技术委员会(SAC/TC 119/SC 1) / Hydrological Instrument Subcommittee of National Hydrological Standardization Technical Committee (SAC/TC 119/SC 1)
Issuing Authority
中华人民共和国国家质量监督检验检疫总局、中国国家标准化管理委员会 / General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China, Standardization Administration of the People's Republic of China

Catalogue

前言 → 1 范围 → 2 规范性引用文件 → 3 术语和定义 → 4 要求 → 5 试验方法 → 6 检验规则 → 7 标志、使用说明书 → 8 包装、运输、贮存 Foreword → 1 Scope → 2 Normative References → 3 Terms and Definitions → 4 Requirements → 5 Test Methods → 6 Inspection Rules → 7 Marking and Instruction Manual → 8 Packaging, Transport, and Storage

Scope

This standard specifies the terms and definitions, classification, technical requirements, test methods, inspection rules, and marking, packaging, transport, and storage requirements for sediment measurement and particle size analysis instruments in open channels (hereinafter referred to as instruments). This standard applies to instruments for sediment measurement and particle size analysis in rivers, lakes, reservoirs, canals, and other water bodies.

本标准规定了河流泥沙测验及颗粒分析仪器(以下简称仪器)的术语和定义、分类、技术要求、试验方法、检验规则及标志、包装、运输、贮存等要求。本标准适用于江河、湖泊、水库、渠道等泥沙测验及颗粒分析的仪器。

Normative References

GB/T 6005 GB/T 9359-2001 GB/T 18185-2000 GB/T 18522.6 GB/T 19677 GB/T 50095 SL 07-2006

Keywords

河流泥沙测验 (river sediment measurement) 颗粒分析 (particle size analysis) 仪器 (instrument) 技术条件 (technical requirements) 悬移质 (suspended load) 推移质 (bed load) 床沙 (bed material) 含沙量 (sediment concentration)

Application Summary AI generated

Water resources departments, hydrological agencies, research institutes, and instrument manufacturers use this standard to design, produce, and test sediment measurement and particle size analysis instruments. It ensures accuracy and reliability in measuring suspended load, bed load, and bed material, which is critical for river management, reservoir operation, and sediment research.

水利部门、水文机构、科研院所和仪器制造商使用该标准来设计、生产和检验河流泥沙测验及颗粒分析仪器。该标准确保仪器在测量悬移质、推移质和床沙时的准确性和可靠性,对河流治理、水库调度和泥沙研究至关重要。

AI Summary AI generated

This standard specifies fundamental technical requirements for sediment measurement and particle size analysis instruments in open channels, covering terms, classification, technical requirements, test methods, inspection rules, and marking/packaging. It applies to instruments used in rivers, lakes, reservoirs, and canals. The standard details technical parameters and performance requirements for suspended load, bed load, and bed material samplers, as well as sediment concentration meters and particle size analyzers, including sampling efficiency, measurement error, and environmental adaptability, along with corresponding test methods. It provides a unified basis for instrument design, production, and quality inspection.

该标准规定了河流泥沙测验及颗粒分析仪器的基本技术条件,包括术语定义、分类、技术要求、试验方法、检验规则及标志包装等。适用于江河、湖泊、水库、渠道等水体的泥沙测验。标准详细规定了悬移质、推移质、床沙采样器及含沙量测量仪、粒度分析仪的技术参数和性能要求,如采样效率、测量误差、环境适应性等,并提供了相应的试验方法。该标准为仪器设计、生产和质量检验提供了统一依据。

Key Sentences extracted from text

1.

本标准规定了河流泥沙测验及颗粒分析仪器(以下简称仪器)的术语和定义、分类、技术要求、试验方法、检验规则及标志、包装、运输、贮存等要求。

2.

悬移质泥沙采样器的技术参数按SL 07—2006中第5章的规定执行。

3.

进口流速应与天然流速接近,其口门平均进口流速系数宜在0.95~1.15之间。

4.

在建立率定工作模型时,其含沙量随机误差(相对误差统计标准差)应不大于5%,系统偏差(相对误差均值)应不大于1%。

5.

仪器在包装状态下,应能承受GB/T 9359--2001中所规定的振动和自由跌落等试验。

7.

This standard specifies the terms and definitions, classification, technical requirements, test methods, inspection rules, and marking, packaging, transport, and storage requirements for sediment measurement and particle size analysis instruments in open channels (hereinafter referred to as instruments).

8.

The technical parameters of suspended sediment samplers shall comply with the provisions in Chapter 5 of SL 07-2006.

9.

The inlet velocity should be close to the natural velocity, and the average inlet velocity coefficient at the mouth should be between 0.95 and 1.15.

10.

When establishing the calibration working model, the random error of sediment concentration (relative error statistical standard deviation) shall not exceed 5%, and the systematic deviation (mean relative error) shall not exceed 1%.

11.

The instrument in the packaged state shall be able to withstand vibration and free fall tests as specified in GB/T 9359-2001.

Related Standards

Transparency note: The application summary and key sentences on this page were automatically generated by AI from the standard's original text. This content has not been human-verified and should not be used for compliance or regulatory purposes. Always refer to the official standard document from the issuing authority.