GB/T 16886.23-2023
ActiveBiological evaluation of medical devices—Part 23: Tests for irritation
医疗器械生物学评价 第23部分:刺激试验
Application Summary AI generated
This standard specifies test methods for evaluating the potential of medical devices to cause irritation to skin, mucous membranes, or other tissues. It is applied in the biological safety assessment of medical devices, including surgical instruments, implants, and wound dressings, to ensure they do not produce harmful local reactions during clinical use. The standard is used by manufacturers and testing laboratories for preclinical evaluation and regulatory compliance in China.
Related Standards
GB/T 19634-2005
In vitro diagnostic test systems-General technical requirements for blood-glucose monitoring systems for self-testing
GB/T 47486-2026
General technical requirements for vasculature-on-a-chip
GB/Z 30154-2013
Medical laboratories―Guidance on laboratory implementation of GB/T 22576-2008
GB/Z 16886.22-2022
Biological evaluation of medical devices—Part 22:Guidance on nanomaterials
GB/Z 43280-2023
Medical laboratories—Practical guidance for the estimation of measurement uncertainty
GB/Z 43281-2023
Guidance for supervisors and operators of point-of-care testing(POCT) devices
GB/Z 44363-2024
Pyrogenicity—Principles and methods for pyrogen testing of medical devices
GB/Z 136-2026
Medical laboratories—Reagents for staining biological specimen—Guidance for users
Transparency note: The application summary and key sentences on this page were automatically generated by AI from the standard's original text. This content has not been human-verified and should not be used for compliance or regulatory purposes. Always refer to the official standard document from the issuing authority.