GB/T 16886.17-2005
ActiveBiological evaluation of medical devices -- Part 17: Establishment of allowable limits for leachable substances
医疗器械生物学评价 第17部分:可沥滤物允许限量的建立
Application Summary AI generated
This standard provides a framework for determining the maximum acceptable levels of chemical substances that can leach from medical devices into the body. It is applied during the biological safety evaluation of medical devices, specifically for manufacturers and testing laboratories to establish safe exposure limits for leachable compounds based on toxicological risk assessment. The standard is critical for ensuring patient safety in devices that have prolonged or permanent contact with tissues or bodily fluids.
Related Standards
GB/T 19634-2005
In vitro diagnostic test systems-General technical requirements for blood-glucose monitoring systems for self-testing
GB/T 47486-2026
General technical requirements for vasculature-on-a-chip
GB/Z 30154-2013
Medical laboratories―Guidance on laboratory implementation of GB/T 22576-2008
GB/Z 16886.22-2022
Biological evaluation of medical devices—Part 22:Guidance on nanomaterials
GB/Z 43280-2023
Medical laboratories—Practical guidance for the estimation of measurement uncertainty
GB/Z 43281-2023
Guidance for supervisors and operators of point-of-care testing(POCT) devices
GB/Z 44363-2024
Pyrogenicity—Principles and methods for pyrogen testing of medical devices
GB/Z 157-2026
In vitro diagnostic test systems—Requirements and recommendations for detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) by nucleic acid amplification methods
Transparency note: The application summary and key sentences on this page were automatically generated by AI from the standard's original text. This content has not been human-verified and should not be used for compliance or regulatory purposes. Always refer to the official standard document from the issuing authority.